HTM 01 05 audit for dental practices.

An HTM 01 05 audit is a regular check of how your practice cleans, sterilises, stores and traces instruments. This free audit has 41 checks in 6 sections, scores each one, and turns every gap into an action with an owner and a date.

At a glance

Dental Decontamination (HTM 01-05) Audit, in short.

QuarterlyHow often

Run it quarterly, whole service.

41The checks

6 sections, each scored, with a space for findings and actions.

Whole siteWho runs it

Decontamination lead or practice manager. Evidence is the steriliser and washer logs, validation certificates, load records, training files, and the decontamination area itself.

FreeCost

Free to run on every plan. No card needed.

What it covers

What the HTM 01 05 audit checks

You audit the whole practice in one visit, not one person or room. Walk the decontamination area with the steriliser logs, validation certificates and load records to hand. The questions follow the instrument from dirty to clean to stored. Use it as a dental infection control audit that focuses on dental decontamination. It is not a copy of the guidance and not a blank form. Your practice procedure and the guidance itself remain the standard.

Decontamination area layout and flow

Instruments are decontaminated in a dedicated room or a clearly separated area, not in a clinical treatment space while patients are present.

Cleaning and inspection of instruments

How are used instruments cleaned?

Washer-disinfectors and sterilisers

Every steriliser and washer-disinfector has a record of installation and validation, and the last validation is in date.

Packaging, storage and traceability

Sterilised instruments are packaged or stored in line with the practice procedure, and pouches are intact and dry.

Water quality and waste

A current water safety or Legionella risk assessment covers the dental unit water lines and is in date.

Training, governance and actions

Everyone who decontaminates instruments has recorded training and a competence check for the task.

How to use it

How to run the Dental Decontamination (HTM 01-05) Audit, step by step.

01Create a free account

Register for a free CareStream account. No card is needed and every audit is free to run.

02Pick the audit

Open Audits and choose Dental Decontamination (HTM 01-05) Audit. It covers the whole practice, so there is no person or room to pick.

03Work through the six sections

Answer each check in the decontamination area, using the logs and records. Add a finding wherever something is missing.

04Review the score and set actions

See the score for each section, then give every gap an owner and a due date.

05Repeat each quarter

Run the HTM 01 05 audit every quarter and compare results. Your scores build into a trend.

Train the team

Free CPD training that supports this audit.

When an audit finds a gap, the fix is often training. These courses are free, so you can assign them straight away.

Infection Prevention and Control

Hand hygiene, PPE, cleaning, laundry and waste. CPD certified, and free for every member of staff on every plan.

See the free course

The full question set

All 41 checks, section by section.

You can read every check before you sign up. Inside CareStream each one is scored, can carry a finding and can raise an action.

Decontamination area layout and flow (7)

1. Instruments are decontaminated in a dedicated room or a clearly separated area, not in a clinical treatment space while patients are present.

2. The layout lets used instruments move one way, from dirty to clean, without crossing back.

3. Dirty and clean zones are clearly marked and the sink, washer-disinfector and steriliser are placed to support that flow.

4. Describe what you saw that breaks the dirty to clean flow and what has been agreed to fix it.

5. Floors, work surfaces, sinks and cabinets in the area are intact, smooth and visibly clean.

6. Hand wash basins, soap, paper towels and a pedal bin are available in or next to the area.

7. Only trained staff were using the decontamination area during the visit.

Cleaning and inspection of instruments (7)

1. How are used instruments cleaned?

2. Where manual cleaning is used, the written method names the water temperature, detergent and brush and is displayed.

3. Used instruments are kept moist or in a covered container between treatment and cleaning, and not left in the treatment room unprotected.

4. Instruments are inspected under a light and magnifier after cleaning, and any with residue or damage are removed or recleaned.

5. Staff wear the protective clothing, gloves and eye protection set out in the practice procedure when handling used instruments.

6. Single use items are discarded and not put through the decontamination cycle.

7. Number of instrument sets or trays checked on the day that showed visible residue or damage.

Washer-disinfectors and sterilisers (8)

1. Every steriliser and washer-disinfector has a record of installation and validation, and the last validation is in date.

2. Give the reason validation is missing or out of date and the date it will be done.

3. Date of the most recent annual validation of the steriliser or sterilisers.

4. Periodic tests (for example daily, weekly) set out in the practice procedure were done and logged on every working day checked.

5. Each steriliser cycle has a printout or electronic record showing it reached the required temperature, pressure and time.

6. The decontamination equipment is serviced as planned and faults are logged with action taken.

7. Written action is recorded when a test failed or a cycle was out of range, including removing the load from use.

8. Water used in the steriliser and washer-disinfector is of the type the manufacturer specifies and the supply is logged.

Packaging, storage and traceability (7)

1. Sterilised instruments are packaged or stored in line with the practice procedure, and pouches are intact and dry.

2. Each pouch or tray is labelled with a sterilisation date or batch and a use by date.

3. Number of out of date or damp pouches found in drawers or storage during the check.

4. Clean instruments are stored in closed, clean, dry cupboards or drawers away from sinks and dirty areas.

5. A record links each steriliser load to the patients treated with those instruments.

6. Pick one patient from last week and trace the instruments used back to the load record.

7. Explain what stopped the trace and who will fix the record process.

Water quality and waste (4)

1. A current water safety or Legionella risk assessment covers the dental unit water lines and is in date.

2. Dental unit water lines are flushed and disinfected as the written procedure says, and records are kept.

3. Sharps and clinical waste from the decontamination area go into the right containers, which are not overfilled.

4. Date of the last water test result reviewed by the practice.

Training, governance and actions (8)

1. Everyone who decontaminates instruments has recorded training and a competence check for the task.

2. Give names of staff missing a competence check and the date each will be done.

3. Staff immunisation status is recorded for those handling used instruments.

4. A named decontamination lead is recorded and staff can say who it is.

5. A written decontamination policy and procedures are in date and match what staff actually do.

6. The findings of the previous audit were acted on and closed by the agreed dates.

7. Overall rating of how well decontamination is managed.

8. List the gaps found, who owns each action and the date it is due.

Who it is for

Care services this audit suits.

Care services
  • Dental practices
What it evidences

Evidence is the steriliser and washer logs, validation certificates, load records, training files, and the decontamination area itself.

Regulations it supports
  • Health and Social Care Act 2008 (Regulated Activities) Regulations 2014, regulation 12(2)(h): Safe care and treatment: assessing the risk of, and preventing, detecting and controlling the spread of, infections
  • Health and Social Care Act 2008 (Regulated Activities) Regulations 2014, regulation 15: Premises and equipment
  • HTM 01-05: Decontamination in primary care dental practices

This audit checks that things are in place, recorded and reviewed. It does not give clinical or legal advice.

FAQ

Frequently asked questions.

What is an HTM 01 05 audit?

It is a check of how well your practice follows its decontamination procedures for dental instruments. You look at layout, cleaning, washers and sterilisers, packaging, storage, traceability, water quality and staff training, then record gaps as actions.

What is HTM 01 05?

HTM 01 05 is NHS England guidance called Decontamination in primary care dental practices. This audit is built around the same topics, but it does not quote the guidance. Read the guidance itself for the detail.

How often should a dental decontamination audit be done?

We set this audit to quarterly. Your own infection prevention policy decides how often you audit, so follow it if it says otherwise.

Who should carry out the audit?

Usually the decontamination lead or the practice manager. They should know the decontamination area and be able to read the steriliser and washer logs.

Is this a dental infection control checklist?

It is a dental infection control checklist for the decontamination side of infection control. For wider checks such as hand hygiene or sharps, use the related audits below.

Which regulations and standards does it support?

It helps evidence infection control under Regulation 12(2)(h) and premises and equipment under Regulation 15 of the Health and Social Care Act 2008 (Regulated Activities) Regulations 2014. It also follows the topics of HTM 01 05.

Does the audit cover sterilisation records?

Yes. It checks validation dates, periodic test logs, cycle records, faults and failed loads. Together these make a decontamination audit of your dental sterilisation process.

Is the audit really free?

Yes. Every audit in the CareStream library is free to run on a free account, with scoring, findings, actions and a report. Copying and editing an audit is part of the Enterprise plan.

Run the Dental Decontamination (HTM 01-05) Audit free.

Create a free account and run it today. No card needed.

More free audits: Infection Control Audit, Hand Hygiene Observation Audit, Clinical Equipment, Servicing and Validation Audit.

Accreditations and compliance

  • Good Business Charter accredited
  • CPD Certified training, The CPD Certification Service
  • Registered with the Information Commissioner's Office
  • Disability Confident Committed
  • GDPR compliant

CPD Provider No. 50224 · ICO registration ZC221613