PUWER checklist for clinical equipment.
This PUWER checklist is a free quarterly audit of the clinical equipment in your practice or clinic. It takes you through 43 checks in 7 sections, scores each one, and turns every gap into an action with an owner and a date. It covers the register, servicing, calibration, faults and training.
Clinical Equipment, Servicing and Validation Audit, in short.
Run it quarterly, whole service.
7 sections, each scored, with a space for findings and actions.
Practice manager, clinic manager or equipment lead. It evidences the equipment register, service and calibration records, validation logs, PAT labels, fault log and staff training records.
Free to run on every plan. No card needed.
What the clinical equipment audit checks
You audit the whole service, not one person. Walk the premises with your equipment register and check items against it. The questions then follow each item through servicing, calibration, testing, faults, training, cleaning and storage. It ends with a review of actions. This is a completed check, not a blank PUWER inspection form or a PUWER checklist template. It draws on PUWER regulations 5, 6 and 9, but it does not cover every part of PUWER.
A written register lists every item of clinical equipment used for care or treatment.
A planned servicing schedule covers every item that needs servicing, with due dates.
Equipment that needs calibration (for example scales, blood pressure monitors, thermometers, gauges) is listed with calibration due dates.
A fault reporting route exists and staff can say how to use it.
Staff who use each higher risk item have a recorded induction on that item.
Each item of equipment has a cleaning routine and records show it is followed.
The equipment register was reviewed by a named person within the last 3 months.
How to run the Clinical Equipment, Servicing and Validation Audit, step by step.
Register for a free CareStream account. No card is needed and every audit is free to run.
Open Audits and choose Clinical Equipment, Servicing and Validation Audit. It covers your whole service, so there is no person to select.
Check the register, service records, calibration labels and fault log, and spot check items on the floor. Add a finding wherever something is missing.
See the score for each section, then give every gap an owner and a due date.
We set this audit to quarterly. Run the PUWER checklist again each quarter so you can see progress in your results.
Free CPD training that supports this audit.
When an audit finds a gap, the fix is often training. These courses are free, so you can assign them straight away.
Everyday health and safety, fire safety and reporting hazards. CPD certified, and free for every member of staff on every plan.
All 43 checks, section by section.
You can read every check before you sign up. Inside CareStream each one is scored, can carry a finding and can raise an action.
Equipment register (7)
1. A written register lists every item of clinical equipment used for care or treatment.
2. List the equipment found in use that is missing from the register.
3. Each register entry shows make, model, serial or asset number and location.
4. Each register entry names the person responsible for servicing or checking it.
5. Items that have been retired or disposed of are marked as removed from the register with a date.
6. Number of items spot checked from the floor against the register
7. Number of those items that matched the register entry
Servicing and planned maintenance (7)
1. A planned servicing schedule covers every item that needs servicing, with due dates.
2. List the items with no servicing schedule entry.
3. Number of items on the register with a service due date that has passed
4. Servicing was done by the manufacturer or by a person or company the manufacturer or provider has judged competent.
5. Each service record shows the date, who did the work, what was done and the next due date.
6. Service reports with advice or faults found have a recorded response from a named person.
7. Where a service contract is in place, how current is it?
Calibration, testing and validation (8)
1. Equipment that needs calibration (for example scales, blood pressure monitors, thermometers, gauges) is listed with calibration due dates.
2. Number of calibrated items found past their calibration due date
3. Calibration certificates or labels are kept and show the date and the standard used.
4. Does the service use sterilisers, washer disinfectors or similar decontamination equipment?
5. Validation and periodic test records are in date for each steriliser and washer disinfector.
6. Daily or weekly test results for that equipment are logged, signed and reviewed by a named person.
7. Portable appliance testing (PAT) records show every electrical item tested within the interval set by the practice risk assessment.
8. List the electrical items with no test label or an overdue test.
Faults, repairs and removal from use (7)
1. A fault reporting route exists and staff can say how to use it.
2. Number of faults reported in the last 3 months
3. Each reported fault has a recorded action and a date it was closed.
4. Faulty equipment is taken out of use, labelled and kept apart from working stock.
5. Describe the faulty items found still available for use.
6. Repaired equipment is checked and recorded as safe before it goes back into use.
7. Safety alerts and recalls from manufacturers or the regulator were checked and the response recorded.
Training and safe use (5)
1. Staff who use each higher risk item have a recorded induction on that item.
2. Number of staff records checked for equipment training
3. Number of those staff with equipment training recorded as in date
4. Manufacturer instructions are available to staff next to or near the equipment.
5. Single use items are not being reused and the reprocessing rules for reusable items are followed.
Cleaning, storage and environment (4)
1. Each item of equipment has a cleaning routine and records show it is followed.
2. Rate the general condition and cleanliness of the equipment seen (1 poor, 5 very good).
3. Equipment is stored where it is protected from damage, dust and misuse.
4. Emergency equipment (for example oxygen and a defibrillator) is checked on the schedule set and the checks are recorded.
Review and actions (5)
1. The equipment register was reviewed by a named person within the last 3 months.
2. Findings from the last equipment audit were acted on and the actions are signed off.
3. The provider has a written equipment risk assessment or policy that was reviewed within the last year.
4. Overall assessment of equipment safety at this visit
5. Record the main actions, who owns each and the due date.
Care services this audit suits.
Care services- Dental practices
- Independent hospitals
- Clinics and independent doctors
- GP practices
It evidences the equipment register, service and calibration records, validation logs, PAT labels, fault log and staff training records.
- Reg 15: Premises and equipment (Health and Social Care Act 2008 (Regulated Activities) Regulations 2014)
- Reg 12(2)(e): Safe care and treatment: equipment used for care or treatment is safe for such use and used in a safe way (Regulated Activities Regulations 2014)
- PUWER reg 5: Provision and Use of Work Equipment Regulations 1998: Maintenance
- PUWER reg 6: Provision and Use of Work Equipment Regulations 1998: Inspection
- PUWER reg 9: Provision and Use of Work Equipment Regulations 1998: Training
This audit checks that things are in place, recorded and reviewed. It does not give clinical or legal advice.
Frequently asked questions.
What is a PUWER checklist?
PUWER is the Provision and Use of Work Equipment Regulations 1998. A PUWER checklist is a list of checks that equipment is maintained, inspected and used by trained staff. This audit applies that thinking to clinical equipment.
What is a PUWER inspection checklist for a clinic?
A PUWER inspection checklist for a clinic looks at each item of equipment: is it on the register, serviced, tested, safe to use and used by trained staff. This audit asks those questions across the whole service rather than one machine.
How often should the audit be run?
We set this audit to quarterly. No regulation we cite sets how often you must audit, so your own policy decides. Service, calibration and PAT intervals are set by your risk assessment and the manufacturer, not by this audit.
Who should run it?
Usually the practice manager, clinic manager or equipment lead. They should be able to read service reports and calibration records and check items against the register.
Is this a medical equipment maintenance checklist or an equipment calibration log?
It is a check of the records you already hold. It is not a blank medical equipment maintenance checklist or an equipment calibration log template. It looks at whether calibration labels and service records are in date and kept, and your own logs stay as they are.
Which regulations does it support?
It supports Regulation 15 on premises and equipment and Regulation 12(2)(e) on safe equipment, under the Health and Social Care Act 2008 (Regulated Activities) Regulations 2014. It also draws on PUWER regulations 5, 6 and 9: maintenance, inspection and training.
What counts as a medical equipment audit here?
It covers equipment used for care or treatment, including calibrated items, electrical items and, where you use them, sterilisers and washer disinfectors. It also checks emergency equipment such as oxygen and a defibrillator.
Is the audit really free?
Yes. Every audit in the CareStream library is free to run on a free account, with scoring, findings, actions and a report. Copying and editing an audit is part of the Enterprise plan.
Run the Clinical Equipment, Servicing and Validation Audit free.
Create a free account and run it today. No card needed.